Intra- and Postoperative Variation in Ocular Response Analyzer Parameters in Keratoconic Eyes After Corneal Cross-linking

Journal of Refractive Surgery  Vol. 26   No. 9   September 2010

See the Full text
Paolo Vinciguerra, MD; Elena Albè, MD; Ashraf M. Mahmoud, PhD; Silvia Trazza, BS; Farhad Hafezi, MD and Cynthia J. Roberts, PhD

Share/Save/Bookmark

 

PURPOSE

To analyze intra- and postoperative variation in Ocular Response Analyzer (ORA, Reichert Ophthalmic Instruments) parameters in 24 keratoconic eyes undergoing corneal cross-linking (CXL).

METHODS

In a prospective clinical study, corneal hysteresis (CH), corneal resistance factor (CRF), peak 1 and peak 2 amplitude, corneal-compensated and Goldmann-correlated intraocular pressure (IOP) were evaluated using the ORA. The thinnest cornea point was measured with the Pentacam (Oculus Inc). Corneal topography and endothelial cell count were performed. Measurements were recorded at baseline; intraoperatively after epithelium removal, riboflavin impregnation, and ultraviolet A irradiation; and postoperatively after corneal re-epithelialization and at 1, 6, and 12 months.

RESULTS

A statistically significant reduction of the thinnest cornea point from 462±23.24 µm was observed at the end of the CXL procedure intraoperatively and at 1- and 6-month follow-up (P<.05). A significant increase in the thinnest cornea point to 624±31.72 µm was found after re-epithelialization (P<.05), and no significant changes were observed at 1 year postoperatively. Mean CH and CRF did not change significantly after de-epithelialization, but were noted to significantly increase after CXL intraoperatively and postoperatively at 1-month follow-up. At 6 and 12 months postoperatively, CH and CRF were not statistically significantly different from preoperatively. Peak 1 and peak 2 decreased intraoperatively from 276±52 and 228±47 to 172±42 and 131±42, respectively, at the conclusion of CXL (P<.05), and were noted to increase to 493±41 and 444±51, respectively, at 6-month follow-up. Corneal-compensated IOP and Goldmann-correlated IOP increased at 1 month after CXL (P>.05).

CONCLUSIONS

The results showed a significant change in ORA parameters and the thinnest cornea point during and after the CXL procedure and a high correlation between peak amplitudes and corneal asymmetry, providing insight to the bioelastic and biomechanical behavior of the cornea during and after CXL. [J Refract Surg. 2010;26(9):669-676.]

doi:10.3928/1081597X-20100331-01

ABOUT THE AUTHORS

From the Department Ophthalmology, Istituto Clinico Humanitas, Rozzano, Milan, Italy (Vinciguerra, Albè, Trazza); the Department of Ophthalmology and Biomedical Engineering, The Ohio State University, Columbus, Ohio (Mahmoud, Roberts); and Institute for Refractive and Ophthalmic Surgery, Zurich, Switzerland (Hafezi).

The authors have no proprietary or financial interest in the materials presented herein.

Correspondence: Paolo Vinciguerra, MD, Dept of Ophthalmology, Istituto Clinico Humanitas, Rozzano, Milan, Italy. Tel: 39 02 8224 4680; E-mail: info@vincieye.it

Received: January 22, 2009; Accepted: February 9, 2010

Posted online: April 15, 2010

See the Full text